Patients at risk of harm from ‘inappropriate placement’ of sats probes

Inappropriate placement can produce a reading up to 50% lower or 30% higher than the real value. 


18 December 2018
person in hospital bed

Adobe

‘Inappropriate placement’ of pulse oximeter probes can lead to staff being falsely reassured about a patient’s condition.

An NHS Improvement Patient Safety Alert has warned that patients may be at risk of harm through inappropriate placement of pulse oximeter probes.



Measurement of oxygen saturation, using a pulse oximeter probe, is routinely undertaken as part of patients’ vital signs during diagnosis and ongoing monitoring.

Oximeter probes can be single or multiple use and are designed to attach to specific parts of the body. Adult oximeter probes can be attached to either a finger or an ear, but are not interchangeable between these sites, whilst probes for babies and children need to be selected according to the patient’s weight.

The alert wants that if an oximeter probe intended for the finger is attached to the ear (or vice versa), or a probe intended for an adult is attached to a baby or a child (or vice versa), it can produce a reading up to 50% lower or 30% higher than the real value.

‘Falsely reassured’.

Warning that that staff may be falsely reassured about a patient’s condition, when in reality the patient is deteriorating, or may make an inappropriate intervention when in fact a patient is stable or improving.


The safety alert asks NHS organisations to ensure staff have access to appropriate equipment and the information they need to use these devices safely.

Research completed by NHS Improvement suggested that a substantial proportion of staff do not know that finger probes can give misleading results if attached to ears and that a quarter of staff do not have routine access to probes specifically for the ear – even though in almost all clinical settings some patients will need these.

The alert calls on NHS organisations to ensure staff have access to the equipment they need and education them on the risks of misplacement.




Popular

Close-up of rapid Covid-19 home lateral flow antigen test with positive result

Special paid leave for Covid sickness and isolation scrapped for NHS workers

1 July 2022

student nurse staff nurse

Student nurses ‘used and abused’ on placements

13 June 2022

Patient lying on bed while nurse writing on clipboard in hospital

Nurses told to stop ‘overworking’ by staying late without pay

27 June 2022

Insight

Busy A&E waiting room

‘The NHS is having its worst winter ever – and the reasons run much deeper than COVID’

28 January 2022

Hospital curtain intensive care

‘During the Downing Street Christmas Party we were caring for dying patients and forbidden from seeing family’

8 December 2021

Vaccine inPPE

‘Making vaccination compulsory for NHS frontline workers likely to make patients suffer’

19 November 2021


Related Posts

neodymium magnets squares, black background
Clinical

Warning issued around the ingestion of ‘super strong’ neodymium magnets often found in toys

21 May 2021
Shutterstock
Clinical

New guidance issued on managing allergic reactions following COVID-19 vaccination

10 December 2020
Pleural Effusion
Clinical

Patient safety warning issued over the rapid drainage of pleural effusion fluid from chest drains

3 December 2020
NursingNotes

© 2019 NursingNotes.co.uk

Navigate Site

  • Who are we?
  • Terms of Service
  • Privacy Policy
  • Contact Us
  • Agenda for Change Pay Scales

Follow Us

No Result
View All Result
  • Latest News
  • Clinical
  • Education
  • Health Politics
  • Opinion
  • Resources

© 2019 NursingNotes.co.uk